NOTE: Do not empty or fill the tank with high-temperature differences as the hot or 

cold can damage the machine. 

 

 

 

 

 

 

DIFFERENT WAYS OF CLEANING 

 

Gennral Cleaning

use only water to clean under the temperature of about 50

Enhanced Cleaning

add few drops of standard cleaning solutions, liquid soap, or 

detergent, or any other non-acidic cleaning agents.

 

Extensive Cleaning--

removing tarnish, carbon & rust from non-plated metals, it is 

recommended to use specific cleaning solution associated with ultrasonic cleaners.

 

 

 

 
 
 
The cleaning solution will deteriorate in effectiveness over time and use. It is important 
to regularly change the fluid and carefully wash the inside of the cleaning tank in order 
to preserve the effectiveness and longevity of the cleaner.    Do not use corrosive or 
abrasive cleaning tools to clean the tank which must be wiped down and dried before it 
can be re-connected to the electrical supply.   
 
 
NOTE: If the machine starts to spark, smoke, smell or displays any other fault the 
operator must immediately stop the machine, and isolate it from the electrical supply 
and contact the supplier.   
 
 
 
 
 
 
 
 

WARNING

 : Strong acid or alkaline cleaning solution will cause corrosion, rust and 

even puncture of tank or machine body. To solve this problem, please dilute the solution 
to mild PH   

Summary of Contents for SR236

Page 1: ...nless steel construction Industrial grade integrated circuit 0 80 digital temperature control 1 99 minutes digital timer Degassing function 28KHz or 40KHz industrial grade transducers ULTRASONIC CLEAN...

Page 2: ...all controls are set to off and the drain valve is closed 3 Ensure that the power cord is securely plugged into the cleaner bottom panel Ultrasonic cleaning is achieved by the mechanical action of ca...

Page 3: ...may damage the transducers Always place items gently into the tank and use the basket whenever possible 4 The more items you place into the cleaning bath the less cleaning efficient you can get 5 Do...

Page 4: ...hed fluid must be dried immediately 16 In case of emergency or failure to follow the aforementioned items disconnect the main supply by removing the plug from the main socket APPLICATIONS The uses for...

Page 5: ...cleaner continuously for more than one hour 3 Temperature setting Press temperature temperature quickly will increase decrease the temperature setting 1 at a time press and hold will increase decrease...

Page 6: ...eriorate in effectiveness over time and use It is important to regularly change the fluid and carefully wash the inside of the cleaning tank in order to preserve the effectiveness and longevity of the...

Page 7: ...cleaning temperature to enhance the cleaning efficiency Have a digital controller of high precision and a long service life Are of high performance Eliminate harsh chemicals Are environment friendly...

Page 8: ...del No SR236 Quantity of transducers 6 pcs Ultrasonic Power 360W Ultrasonic Frequency 40KHz Heating Power 400W Control Mode Digital control Tank Capacity 15L Tank Size mm 330 300 150 L W H Voltage AC...