Dispenser calibration
When a new dispenser has been connected to the system, it needs to be calibrated before it
can be used to deposit fluid or sense the physical height of a surface. From time to time, the
dispenser may need to be recalibrated due to changes in its contents or drift in the system.
The connection of a dispenser to the system is not automatically detected. Open the Calibrate
Dispenser toolbar item to begin the calibration process.
A calibration can be triggered by clicking on the
Calibrate
button. A trace will be displayed as
the dispenser is scanned, followed by a brief progress indicator during the final calibration steps.