current position
button next to the first position in the dialog box. Move the positioner to
the second point (
Top-Right
) and click on the
Use current position
button next to the
second position in the list.
The angle at which the pattern will be rotated, based on these points, is displayed at the bottom
of the dialog box. When you go to print a pattern the software will adjust it to match this rotation.
This rotational calibration will be used for every pattern printed after this point until you clear the
calibration. To clear the calibration, click on the
Clear rotational calibration
button at
the bottom of the dialog box.
Diagnostics
For troubleshooting purposes, the response of the surface sensor can monitored in realtime. To
display the feedback from the sensor, choose the
Window | Diagnostics
menu item. At
regular intervals, the surface sensor will be polled and the results displayed in the diagnostics
window, as shown.
If the dispenser is then moved down until it contacts the surface, the sensor signal will increase
in a stepwise fashion.
51