7
6) The wireless Scanner Application is installed successfully.
Note: if your system is iOS9, need to authorize at Settings - General - Profiles - tap on the Profile - tap
on Trust button.
2.3 STARTING PROBE
2.3.1 Visual inspection
Before and after the ultrasonic visual inspection, check the probe surface or the
fuselage sheath for abnormalities such as peeling, cracks, and bulging.
Waring
Abnormal probes can cause electric shock or injury to people.
Therefore, once any abnormalities are found, you must
immediately stop using the probe and contact Sonostar.
2.3.2 Probe cleaning
Ultrasonic probes should be cleaned and disinfected before and after ultrasound
examination. Please refer to chapter “4 Cleaning and Disinfection”.
Caution
Probes that have not been cleaned or disinfected may cause
bacterial and viral infections
2.3.3 Boot check
Please check the following before diagnosis
1. The probe should not be abnormally heated during use. The probe can be sensed
by hand touching the probe, and if the temperature is significantly higher than
the body temperature (or the probe surface temperature exceeds 40 ° C), the probe
will be stopped.
Caution
If the operator places an abnormally hot probe on the surface
of the patient's skin, it may cause burns.
2. The ultrasound image must not be abnormal after power on, check whether the
functions are normal, including software operation, button function, power, etc.