ISONIC utPod from Sonotron NDT - Operating Manual – Revision 1.18 - Page 30 of 94
5.5.2. Gate Start, Width, Threshold
Click on
/
/
to manipulate
Start
/
Width / Threshold settings
for
Gate A
(or identical buttons for the
Gate B
\
In the
ISONIC utPod
Gate Start
is counted from the surface of the material, which is determined through
keying in
Probe Delay
5.5.3. Draw Gate
Click on
(or identical button for the
Gate B
) then draw
Gate A
(or
Gate B
) in the new desired position on the
A-Scan
.
This way allows redefining all three parameters of the gate
quickly
Refer also to the movie at:
http://www.sonotronndt.com/PDF/OM_utPod/MOV/utPod_Draw_Gate.MOV
Initial Pulse
Display Delay
Time Base
A-Scan
Probe
Delay
Gate Start
Gate Width
Gate