47
Instruction for use | Supercal 5
9.3 Error Messages
If several errors are present at the same time, the individual error mes-
sages are added together and displayed.
If an error occurs for more than one hour, it is stored in the error memory
with date and time (start of error) and duration (in minutes). If an error is
present for less than 60 minutes, it is deleted automatically and without
saving.
Code Error
LCD Reference
Description
Procedure Troubleshooting
1
A/D reference 1
Temperature reference 1 A/D:
A cable of the temperature sensor
is interrupted or not connected.
Check if the sendor probe is connected.
If yes, the sensor probe is defective.
2
A/D reference 2
Temperature reference 2 A/D:
A cable of the temperature sensor
is interrupted or not connected.
Check if the sendor probe is connected.
If yes, the sensor probe is defective.
4
Temperature sensor 1
A/D error
Temperature reference 1 A/D:
A cable of the temperature sensor
is connected but its value can not
be read out.
Cross the probes, if Err2 appears afterwards,
probe 1 is defectiv or it is not the correct
type.
8
Temperature sensor 2
A/D error
Temperature reference 2 A/D:
A cable of the temperature sensor
is connected but its value can not
be read out.
Cross the probes, if Err1 appears afterwards,
probe 1 is defective or it is not the correct
type.
16
Range Sensor 1
Temperature sensor 1 < = min.
Range error
Check the ral value of the temperature to
be measured and the physical installation
of the sensor.
32
Range Sensor 1
Temperature sensor 1 > = max.
Range error
Check the ral value of the temperature to
be measured and the physical installation
of the sensor.
64
Range Sensor 2
Temperature sensor 2 < = min.
Range error
Check the ral value of the temperature to
be measured and the physical installation
of the sensor.
128
Range Sensor 2
Temperature sensor 2 > = max.
Range error
Check the ral value of the temperature to
be measured and the physical installation
of the sensor.
512
Flow Saturation
The flow rate is higher than 1,5 qp
Reduce the flow rate.
1024
OPEN CASE
The SC5 is open
Close the case
2048
Main power cut
Power outage
The main power supply has an issue, please
check the connection.
4096
M1 Power Supply
Error in module 1:
Details must be found into specific
module error
M1 Unsupported
Slot left error
8192
M2 Power Supply
Error in module 2:
Details must be found into specific
module error
M2 Unsupported
Slot right error
9.3.1 Overview of Error Messages
Note: Any combination of errors corresponds to the arithmetic sum of each code. The error
code is displayed on the electronic device's display.