6-14
DNW-7/90/90WS
5.
On the setting menu, set as follows.
PAGE
: VF SETTING
ITEM
: ZEBRA SELECT
→
2
6.
On the setting menu, adjust as follows.
PAGE
: VF SETTING
ITEM
: ZEBRA2 DETECT
Spec.
: Set the condition that zebra pattern
appear at the portion B.
Setting after adjustment
PAGE
: VF SETTING
ITEM
: ZEBRA SELECT
→
1
6-19. Zebra Adjustment
Preparation
.
ZEBRA switch (viewfinder)
→
ON
.
On the setting menu, set as follows.
PAGE
: FUNCTION 1/2
ITEM
: TEST OUT
→
R, G or B
PAGE
: VF SETTING
ITEM
: ZEBRA SELECT
→
1
ITEM
: ZEBRA1 APT
→
MIN
.
OUTPUT/DCC switch (inside panel)
→
CAM/ON
.
Shoot a gray-scale chart in the full underscan’s picture
frame.
Setting point
:
1
Lens IRIS
Spec.
: A = 100
±
2 IRE (NTSC)
A = 700
±
14 mV (PAL)
Adjustment procedure
Test point: TEST OUT connector
1.
On the setting menu, adjust as follows.
PAGE
: VF SETTING
ITEM
: ZEBRA1 DETECT
Spec.
: Set the condition that zebra pattern
appear at the portions A.
2.
On the setting menu, set as follows.
PAGE
: FUNCTION 1/2
ITEM
: TEST SAW
→
ON
3.
On the setting menu, adjust as follows.
PAGE
: VF SETTING
ITEM
: ZEBRA1 APT (Factory setting: 0)
Spec.
: Set the desired width of detection.
4.
On the setting menu, set as follows.
PAGE
: FUNCTION 1/2
ITEM
: TEST SAW
→
OFF
6-19. Zebra Adjustment
A
A
A
B