1-9 (E)
BKP-7090
Adjustment
(1) Make sure that the following adjustments for the MS-37 board are completed, referring to the
BVP-700/700P maintenance manual Vol.1.
n
Read “TEST OUT connector” for “MONITOR OUT connector” in the BVP-700/700P maintenance
manual. Connect the TEST OUT connector to B-ch of a waveform monotor terminated with 75
Z
.
4-3-1.
MS MONITOR GAIN Adjustment
4-3-2.
MS Pedestal Adjustment
4-3-3.
MS Gain Adjustment
(2) Perform the following adjustments, referring to Section 2.
2-1.
AD, DA Systems Adjustment
2-2.
Y REF Level Adjustment
2-3.
R-Y REF Level Adjustment
2-4.
B-Y REF Level Adjustment
Reference File Store
After completing the above adjustments, store the reference file by using the MSU-700. For details, refer
to the system manual available separately.
n
If performing customer-necessary adjustments such as gamma, detail, knee and so on, store the reference
file after their adjustments.
(1) Press the FILE button of the menu operation block. (The FILE button will be lit.)
The File Control Menu appears on the display.
(2) Press
[Reference]
of the File Control Menu.
The Reference File operation menu appears.
(3) Press
[Ref|Store]
of the Reference File operation menu.
The Reference Store area appears.
(4) Press
[Start]
.
When the file storage is completed, a message “Completed” appears.
(5) Press the FILE button to release the menu operation mode.
1-3. Electrical Alignment