Important EMC notices for use in the medical environments
• The HVO-500MD/550MD needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this Instructions for Use.
• The portable and mobile RF communications equipment such as cellular phones can affect the HVO-500MD/550MD.
Warning
The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony
Corporation, may result in increased emissions or decreased immunity of the HVO-500MD/550MD.
WARNING
If the HVO-500MD/550MD should be used adjacent to or stacked with other equipment, it should be observed to verify
normal operation in the configuration in which it will be used.
Guidance and manufacturer’s declaration-electromagnetic emissions
The HVO-500MD/550MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the HVO-500MD/550MD should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The HVO-500MD/550MD uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The HVO-500MD/550MD is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable