1-4.
Optional Accessories
The optional accessory of this unit is as follows:
• Foot Switch FS-24
Note
FS-24 is drip-proof protection class IPx3. Therefore, FS-24 can not be used in the place that is exposed to a liquid like
an operating room. To safety, in case the product is used in the place that is exposed to a liquid like an operating room,
use the product that is compatible with IPx6 or higher.
REMOTE contact point switch connector
Connector specifications (stereo miniature jack)
3
2
1
Pin No.
Signal Name
Function
1
GND
–
2
TRIGER SIGNAL (TTL)
When the condition is changed more than 100 msec, the action is started
according to the setting of the REMOTE contact point switch.
3
BUSY STATUS (TTL)
According to the setting of the REMOTE contact point switch connector,
it becomes high level.
HVO-500MD/HVO-550MD