First line
Second line
Description
Note
Solutions
FACTORY SET-
TINGS
FAILED
Factory setting is
failed
Reset the fac-
tory setting is
failed
_
KEY INHIBIT
*1
Key operation is
inhibited
Key operation
of front panel
is inhibited
Check the System Admin Settings
or the Key Inhibit command is re-
ceived by the system control. (Refer
to Instruction Manual.)
WARN. FAN
*2
A fan rotation
malfunction has
occurred.
System/CPU
fan warning
Restart the unit. When the error is
occurred many times after restart,
replace the fan. (Refer to
)
WARN. TEMP.
*2
Temperature
warning
Abnormal
temperature in
the internal of
the unit
Check the using condition and sat-
isfies the specification.(Refer to
)
WARN. HDD
*2
HDD warning /
abnormal
Abnormal of
internal HDD
condition
Replace the internal HDD.
ERR. BATTERY
Battery error
Battery ran
out
Replace the lithium battery. (Refer
to
*1:
After 3 seconds, automatically disappears.
*2:
Displays it with other indication in turn.
HVO-500MD/HVO-550MD
4-12