99
Error Messages
Chapter
Cha
p
te
r 6
Mi
s
c
el
lane
ous
6
Miscellaneous
Error Messages
Front panel display
Description
Solution
First line
Second line
INITIALIZE ERR.
0000
Startup error
Restart the unit. If the error persists, contact your
local Sony representative.
If the power supply is inadvertently shut off while
the unit is in operation or if you turn off the unit
improperly, the internal HDD may be damaged and
the same error may occur at the next startup. The
unit may start up normally after about 15 minutes in
some cases, but the internal HDD is still in a
damaged state. We recommend replacing the
internal HDD in such cases. Contact your local
Sony representative.
SYSTEM ERR.
0001
Application communication error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0100 to 0199
System fan/temperature error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0200 to 0299
Internal HDD error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0300 to 0399
Database error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0400 to 0499
Recovery failure
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0500 to 0599
File data error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0600 to 0699
Capture board error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0700 to 0799
Optical drive error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0800 to 0899
Audio device error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
0900 to 0999
Front panel error
Restart the unit. If the error persists, contact your
local Sony representative.
SYSTEM ERR.
1000 to 1999
Application error
Restart the unit. If the error persists, contact your
local Sony representative.
MAX REC DATA
Patient ID
Limit for recorded data reached
The limit for the number of recorded data entries
has been reached. Delete any unnecessary
recorded data.