50
English
7
RS-232C interface
(See P68 “8. RS-232C Interface”.)
Reset, peak-hold start, setting/recall of preset values,
setting the comparator value, selecting and outputting
the current value/maximum value/minimum value/
peak-to-peak value.
8
EXT. IN
(See P70 “8-6. Output”.)
External input terminal for outputting data via a RS-232C
connector.
4-3. Function description
4-3-1. Reset key
Display unit
Measuring mode
(Current value, maximum
value, minimum value,
peak-to-peak value)
Preset mode
(
P
and the selected digit
will flash.)
“Error” is displayed
All the numbers for the
channel with an error
flash.
Operation performed when reset
key is pressed
Display is set to zero.
When a preset value has been set
this is recalled.
Preset value is set to zero.
Device is cleared and then returnes
to the measuring state.
Automatically carries out initialization
to calibrate the device to a new
measuring probe. (Before resetting
the measuring probe must be moved
at least 1 mm.)
4-3-2. Preset functions
•
It is possible to set preset values for each of the current,
maximum, and minimum value measuring modes.
•
For instructions on setting the preset value, please refer
to P56, “5-2-1. Setting the preset value.”