Performance Tests
4-8
AWG2021 Service Manual
NOTE
. For operation to specified accuracy, allow the AWG2021 to warm up at
least 20 minutes before doing the performance tests.
H
Load all the files from the Performance Check/Adjustment disk
(063-2171-XX) that comes with this manual into AWG2021 internal
memory. For instructions on loading files, see Loading Files on page 2–14 in
the Instructions for Operation subsection in section 2.
Read Preparation and Conventions on page 4-1. Also, if you are not familiar
with operating the AWG2021, read the subsection Instructions for Operation, in
section 2 before doing any of these procedures.
The following equipment is required to check the performance of the AWG2021.
Table 4-1: Test Equipment
Item Description
Minimum Requirements
Example
Purpose
Precision termination
Impedance: 50Ă
W
, 0.1%
Connectors: BNC
Tektronix Part 011Ć0129ĆXX
Signal termination.
Adapter
Connectors: BNC femaleĆtoĆdual
banana
Tektronix Part 103Ć0090ĆXX
Signal interconnection.
Adapter
Connectors: SMA maleĆtoĆBNC
female
Tektronix Part 015Ć0554ĆXX
Signal interconnection.
BNC dual input (TEE)
adapter
Connectors: BNC
Tektronix Part 103Ć0030ĆXX
Signal interconnection.
BNC cable
(4 required)
Impedance: 50Ă
W
Connectors: BNC
Length: 43 inches
Tektronix Part 012Ć0057ĆXX
Signal interconnection.
Termination board
(Option 03)
Must use example equipment
Tektronix Part 671Ć2957ĆXX
Used to check ECL digital
data output.
ECL Digital Data Out
Cable (Option 03)
Must use example equipment
Tektronix Part 012Ć1408ĆXX
(Standard accessory for Option 03)
Used to check ECL digital
data output.
Test leads
(Option 03)
Must use example equipment
Tektronix Part 012Ć1381ĆXX
Used to check ECL digital
data output.
TTL Digital Data Out
Cable (Option 04)
Must use example equipment
Tektronix Part 174Ć3192ĆXX
(Standard accessory for Option 04)
Used to check TTL digital
data output.
2
×
13 header
(Option 04)
Must use example equipment
Tektronix Part 131Ć3847ĆXX
Used to check TTL digital
data output.
Probe, 10X
(Option 04)
10X probe
Tektronix Part P6139A
Used to check TTL digital
data output.
Test oscilloscope
Bandwidth: >250ĂMHz
Tektronix TDS500 Series Digitizing
Oscilloscope or 2400 Series DigitizĆ
ing Oscilloscope
Checks output signals.
Used in many procedures.
Related Information
Equipment Required
Summary of Contents for Tektronix AWG2021
Page 4: ......
Page 16: ...Service Safety Summary xii AWG2021 Service Manual ...
Page 21: ...Specifications ...
Page 22: ......
Page 39: ...Operating Information ...
Page 40: ......
Page 60: ...Operating Instructions 2 20 AWG2021 Service Manual ...
Page 61: ...Theory of Operation ...
Page 62: ......
Page 66: ...Theory of Operation 3 4 AWG2021 Service Manual ...
Page 67: ...Performance Verification ...
Page 68: ......
Page 123: ...Adjustment Procedures ...
Page 124: ......
Page 135: ...Maintenance ...
Page 136: ......
Page 144: ...Maintenance 6 8 AWG2021 Service Manual ...
Page 188: ...Removal and Installation Procedures 6 52 AWG2021 Service Manual ...
Page 196: ...Troubleshooting 6 60 AWG2021 Service Manual TP100 5V TP101 GND Figure 6 34 Key Board ...
Page 201: ...Options ...
Page 202: ......
Page 222: ...Options and Accessories 7 20 AWG2021 Service Manual ...
Page 223: ...Replaceable Electrical Parts ...
Page 224: ......
Page 226: ...Replaceable Electrical Parts 8 2 AWG2021 Service Manual ...
Page 227: ...Diagrams ...
Page 228: ......
Page 230: ...Diagrams 9 2 AWG2021 Service Manual ...
Page 236: ...9 8 AWG2021 Service Manual ...
Page 237: ...Replaceable Mechanical Parts ...
Page 238: ......
Page 255: ...Replaceable Parts AWG2021 Service Manual 10 17 2 5 1 3 4 A9 2 Figure 10 7 Option 03 ...
Page 257: ...Replaceable Parts AWG2021 Service Manual 10 19 3 1 2 A25 4 Figure 10 8 Option 04 ...
Page 261: ......
Page 262: ......