Sorensen MML Series Operation and Programming Manual
5-48
5.10.15 VOLTAGE Subsystem
:VOLTage
SP
<NRf+>
MIN
MAX
suffix
?
:L2
:L1
SP
SP
<NRf+>
:CURRent
MIN
MAX
suffix
?
SP
:MODE
SP
FAST/1
SLOW/0
?
<NR2>
<NR2>
<NR1>
VOLTage:L1/L2
Type: Channel-specific
Description:
Sets voltage of static load during constant voltage mode.
Syntax: VOLTage:L1
VOLTage:L2
Parameters:
For valid value range, refer to respective specification.
Example:
VOLT:L1 8V
Sets voltage of load L1 as 8V.
VOLT:L2 24V
Sets voltage of load L2 as 24V.
VOLT:L1 MAX
Sets voltage of load L1 as the maximum
value.
VOLT:L2 MIN
Sets voltage of load L2 as the minimum
value.
Query Syntax:
VOLTage:L1?
VOLTage:L2?
VOLTage:L1?
MAX
VOLT:L2?
MIN
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com