Sorensen MML Series Operation and Programming Manual
1-2
•
Automatic fan speed control to reduce noise.
•
Up to 8 channels for one Mainframe.
1.3.2 Load
•
Constant current (CC), constant resistance (CR), and constant voltage (CV)
operation modes.
•
Programmable slew rate, load levels, load periods, and conduct voltage (Von).
•
Programmable dynamic loading with speed up to 20KHz.
•
Minimum input resistance allowing load to sink high current even with low input
voltage (1 V).
•
Selective voltage and current ranges.
•
Remote sensing capability.
•
100 sets of memories to save/recall user-definable setups.
•
10 sets of programs to link files for automatic test.
•
15-bit A/D converter with precision measurement.
•
Short circuit simulation.
•
Automatic GO/NG inspection to confirm UUT within spec.
•
Independent GO/NG signals for each channel.
1.4 Specifications
Mainframe: MML-4
AC input:
115/230 switchable or 100/200 switchable Vac line
Fuse: 5A/250V
Amplitude:
±
10%
Frequency:
47 to 63 Hz
Maximum VA: 180VA
Trigger output: Vlo = 0.8V maximum at Ilo = 1 mA
Vhi = 3.2V minimum at Ihi =
−
40
µ
A
Weight: 24Kg
Dimensions:
Width: 440mm
Height: 177.4 mm (excluding feet), 186mm (including feet)
Depth: 560mm (including Load module)
Load specifications are listed in the tables below.
NOTES
1. All specifications are tested at 20
°
C
∼
30
°
C except otherwise stated.
2. The range of operation temperature is 0
°
C
∼
40
°
C.
3. The specifications of DC current accuracy are tested after the input is applied for 30
seconds.
4. The power of the load module of MML series is supplied from MML-4/MML-2
mainframe.
5. The typical temperature coefficient is 100ppm.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com