Calibration Procedures
7-3
7.3 Required Test Equipment
The required test equipment is shown below.
Type
Required Characteristics
Recommended Model
DMM
5 1/2 digits or more
HP34401A, HP3458A
Current Shunt
0.05% accuracy
PRODIGIT 7550
10 ohms@20mA
VALHALLA 2575A
0.1
ohms@2A
0.01
ohms@20A
0.001
ohms@250A/100A
(250A for models of 63106, 63112)
DC Sources
For:
MML-80-20-102
Use:
DLM80-7.5 and DLM8-75
For:
MML-80-60-301
Use:
DLM80-7.5 and DLM8-75
For:
MML-80-120-601
Use:
DLM80-7.5 and DCS8-125
For:
MML-500-10-301
Use:
DCS600-1.7E and DLM8-75
For:
MML-500-20-601
Use:
DCS600-1.7E and DLM8-75
Oscilloscope 100MHz
Tektronics
TDS340
Extender Cable
32-pin ribbon cable
Chroma part No. 806 31000
DC Power Supply
12V, 3A
GW3030
FAN 100
CFM
FLOWMAX
4715KL-04W-B30(DC:12V)
NOTE
The TPA7, TPB7 ... And VRA5, VRB5 ... are referred to in many of the
troubleshooting/check/adjustment procedures. TPA represents the test point
on side A while TPB that on side B. TPB is only used by the dual channel/
module of the left channel. VRA represents the variable resistor on side A
while VRB that on side B. VRB is only used by the dual channel/module
of the left channel. Side A is isolated from side B.