Sorensen MML Series Operation and Programming Manual
7-10
Press LOAD , wait two seconds, and check the module’s display:
HoFFS
<<
−
offset calibration of high range voltage
XXXX
xxxx: offset data of high range voltage measurement
Press LOAD , wait two seconds, and check the module’s display:
LoFFS
<<
−
offset calibration of low range current
XXXX
xxxx: offset data of low range current measurement
Press LOAD , wait two seconds, and check the module’s display:
HoFFS
<<
−
offset calibration of high range current
XXXX
xxxx: offset data of high range current measurement
Remove input terminals’ short circuit.
2. Voltage measurement full-scale calibration
Press LOAD and the module displays:
2.FS
uL
Set the DC source to 125V/0.1A, and apply to the load module’s input terminals.
Press LOAD and the module displays:
AAAAA
BBBBB
AAAAA represents the load module’s voltage measurement before calibration.
BBBBB represents the load module’s voltage measurement after calibration.
Adjust the knob of the module until DMM (V) reading = BBBBB
±
1 count.
Press LOAD and the module displays:
SAvE
Save the low range full-scale data.