Calibration Procedures
7-11
Press LOAD and the module displays:
3.FS
uH
Set the DC source to 500V/0.1A.
Press LOAD and the module displays:
CCCCC
DDDDD
CCCCC represents the load module’s voltage measurement before calibration.
DDDDD represents the load module’s voltage measurement after calibration.
Adjust the knob of the module until DMM (V) reading = DDDDD
±
1 count.
Press LOAD and the module displays:
SAvE
Save the high range full-scale data.
3. Current measurement calibration
Press LOAD and the module displays:
4.FS
AL
Set the DC source to 5V/2.1A and current shunt range to 2A.
Press LOAD and the module displays:
EEEEE
FFFFF
EEEEE represents the load module’s current measurement before calibration.
FFFFF represents the load module’s current measurement after calibration.
Adjust the knob of the module until shunt current = FFFFF
±
1 count.
Press LOAD and the module displays: