Calibration Procedures
7-13
Adjust the knob of the module until shunt current = 0 to 0.5mA.
Press LOAD and the module displays:
SAvE
Save CCL offset data.
Press LOAD and the module displays:
7.oFF
CCH
Press LOAD and the module displays:
LLLLL
Adjust the knob of the module until shunt current = 0 to 5mA.
Press LOAD and the module displays:
SAvE
Save CCH offset data.
5. Current setting calibration
Set the DC source to 5V/5A and current shunt range to 2A.
Press LOAD and the module displays:
8.FS
CCL
Press LOAD and the module displays:
0.9000
PPPPP
0.9000 represents the load module’s current setting before calibration.
PPPPP represents the load module’s current setting after calibration.