Sorensen MML Series Operation and Programming Manual
7-14
Adjust the knob of the module until shunt current = PPPPP
±
1 count.
Press LOAD and the module displays:
SAvE
Save CCL setting data.
Set the DC source to 5V/10A and current shunt range to 20A.
Press LOAD and the module displays:
9.FS
CCH
Press LOAD and the module displays:
9.0000
RRRRR
9.0000 represents the load module’s current setting before calibration.
RRRRR represents the load module’s current setting after calibration.
Adjust the knob of the module until shunt current = RRRRR
±
1 count.
Press LOAD and the module displays:
SAvE
Save CCH setting data.
6. CR mode calibration
Set the DC source to 20V/6A and current shunt range to 20A.
Press LOAD and the module displays:
10.FS
CrL
Press LOAD and the module displays:
5.0000
SSSSS