Sorensen MML Series Operation and Programming Manual
7-18
3.FS
uH
Set the DC source to 80V/0.1A.
Press LOAD and the module displays:
CCCCC
DDDDD
CCCCC represents the load module’s voltage measurement before calibration.
DDDDD represents the load module’s voltage measurement after calibration.
Adjust the knob of the module until DMM (V) reading = DDDDD
±
1 count.
Press LOAD and the module displays:
SAvE
Save the high range full-scale data.
3. Current measurement calibration
Press LOAD and the module displays:
4.FS
AL
Set the DC source to 5V/15A and current shunt range to 20A.
Press LOAD and the module displays:
EEEEE
FFFFF
EEEEE represents the load module’s current measurement before calibration.
FFFFF represents the load module’s current measurement after calibration.
Adjust the knob of the module until shunt current = FFFFF
±
1 count.
Press LOAD and the module displays: