Sorensen MML Series Operation and Programming Manual
7-20
Press LOAD and the module displays:
SAvE
Save CCL offset data.
Press LOAD and the module displays:
7.oFF
CCH
Press LOAD and the module displays:
LLLLL
Adjust the knob of the module until shunt current = 0 to 30mA.
Press LOAD and the module displays:
SAvE
Save CCH offset data.
5. Current setting calibration
Set the DC source to 5V/15A and current shunt range to 20A.
Press LOAD and the module displays:
8.FS
CCL
Press LOAD and the module displays:
10.800
PPPPP
10.800 represent the load module’s current setting before calibration.
PPPPP represents the load module’s current setting after calibration.
Adjust the knob of the module until shunt current = PPPPP
±
1 count.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com