Calibration Procedures
7-21
Press LOAD and the module displays:
SAvE
Save CCL setting data.
Set the DC source to 5V/125A and current shunt range to 250A.
Press LOAD and the module displays:
9.FS
CCH
Press LOAD and the module displays:
108.00
RRRRR
108.00 represents the load module’s current setting before calibration.
RRRRR represents the load module’s current setting after calibration
Adjust the knob of the module until shunt current = RRRRR
±
1 count.
Press LOAD and the module displays:
SAvE
Save CCH setting data.
6. CR mode calibration
Set the DC source to 1.5V/60A and current shunt range to 100A.
Press LOAD and the module displays:
10.FS
CrL
Press LOAD and the module displays:
0.0500
SSSSS
0.0500 represents the load module’s resistance setting before calibration.