Sorensen MML Series Operation and Programming Manual
7-26
Press LOAD and the module displays:
SAvE
Save CCL measurement data.
Press LOAD and the module displays:
5.FS
AH
Set the DC source to 5V/21A and current shunt range to 20A.
Press LOAD and the module displays:
GGGGG
HHHHH
GGGGG represents the load module’s current measurement before calibration.
HHHHH represents the load module’s current measurement after calibration.
Adjust the knob of the module until shunt current = HHHHH
±
1 count.
Press LOAD and the module displays:
SAvE
Save CCH measurement data.
3. Current setting offset calibration
Set the DC source to 5V/20mA and current shunt range to 20mA.
Press LOAD and the module displays:
6.oFF
CCL
Press LOAD and the module displays:
KKKKK