Calibration Procedures
7-27
Adjust the knob of the module until shunt current = 0 to 1mA.
Press LOAD and the module displays:
SAvE
Save setting offset data.
Press LOAD and the module displays:
7.oFF
CCH
Press LOAD and the module displays:
LLLLL
Adjust the knob of the module until shunt current = 0 to 5mA.
Press LOAD and the module displays:
SAvE
Save CCH offset data.
4. Current setting calibration
Set the DC source to 5V/2A and current shunt range to 2A.
Press LOAD and the module displays:
8.FS
CCL
Press LOAD and the module displays:
1.8000
PPPPP