Calibration Procedures
7-29
Press LOAD and the module displays:
2.5000
SSSSS
2.5000 represents the load module’s resistance setting before calibration.
SSSSS represents the load module’s resistance setting after calibration.
Adjust the knob of the module until DMM (V) reading/shunt current = SSSSS (ohms)
±
1 count.
Press LOAD and the module displays:
SAvE
Save CRL setting data.
Press LOAD and the module displays:
11.FS
CrH
Set the DC source to 50V/2A and current shunt range to 2A.
Press LOAD and the module displays:
10.000
TTTTT
10.000 represents the load module’s resistance setting before calibration.
TTTTT represents the load module’s resistance setting after calibration.
Note: Because of the module’s display limitation, the resistance setting is 100 ohms, not
10 ohms. Adjust the knob of the module until DMM (V) reading/shunt current = TTTTT
(ohms)
±
1 count.
Press LOAD and the module displays:
SavE
Save CRH setting data.