Sorensen MML Series Operation and Programming Manual
7-30
7. CV mode calibration
Set the DC source to 500V/0.1A.
Press LOAD and the module displays:
12.FS
Cu
Press LOAD and the module displays:
475.00
UUUUU
475.00 represents the load module’s CV voltage setting before calibration.
UUUUU represents the load module’s CV voltage setting after calibration.
Adjust the knob of the module until DMM (V) reading = UUUUU
±
1 count.
Press LOAD and the module displays:
SAvE
Save CV setting data.
Recycle the power switch after calibration.