Sorensen MML Series Operation and Programming Manual
3-6
Slew Rate (Rise, Fall A/
µ
S or mA/
µ
S)
Slew rate determines the rate at which the current input of a module changes to a newly
programmed vlaue. There are two slew rate values. One is for rise rate, and another
for fall rate.
Voltage Ranges (Low, High)
There are two voltage ranges for voltage measurement and Von voltage setting. The
low range provides better resolution at low voltage measurements. If any value is over
the maximum of low range, you must select the high range. The voltage range
selection of CC mode is in configuration setting.
3.5.2 Constant Resistance Mode
Figure 3-5 Constant Resistance Mode
In CR mode, the Load will sink a current linearly propotional to the input voltage in
accordance with the programmed resistance. There is a double pole RC filter of input
voltage, so high frequency parts will be removed. The time constant of low pass filter
is about 4.7 mS.
Voltage Ranges (Low, High)
Resistance can be programmed in either of the two ranges, low range and high range.
The low range is used for input voltage in low voltage range while the high range for
input voltage over low voltage range. The current range of CR mode is high range.