Sorensen MML Series Operation and Programming Manual
4-4
Setting CC Values
There are four modes for CC operation: CCL, CCH, CCDL, CCDH. The current levels
are programmed in Amps. The slew rate levels are programmed in milliamps/
µ
S at
low range and in Amps/
µ
S at high range. The timings are programmed in milliseconds.
The setting buffers of four CC modes are independent. Changing the operation range
does not affect the settings of other ranges.
The following examples illustrate how to set the CC values of Load module for model
MML-80-60-301. Before observing the examples, select channel first.
1. Select
Range/Function
Press MODE, use the
^
or
v
key to select CCL, then press ENTER.
CCL: static low range
CCDL: dynamic low range
CCH: static high range
CCDH: dynamic high range
MODE SELECT
CCL
2. Set Current Level
There are 4000 discrete steps from 0 to full scale in each range. Set level1 (A)
current level to 2 amps by pressing 2, ENTER. Set level2 (B) current level to 1
amp by pressing 1, ENTER.
CCL1: 1.9995A
CCL2: 0.9990A
3. Set Slew Rate
There are 250 discrete steps in each range. Set the rise 50 mA/
µ
S and fall slew
rates to 50 mA/
µ
S by pressing 5, 0, ENTER for rise and 6, 0, ENTER for fall slew
rate.
CCL
: 50mA/
µ
S
CCL
: 60mA/
µ
S