Local Operation
4-5
4. Set DYNAmic Function Periods
Dynamic function has period T1 and T2 to be set. Set dynamic period 1 to 0.1 mS,
period 2 to 0.2 mS by pressing 0, ., 1, ENTER and 0, ., 2, ENTER. The range of
Dynamic period is from 0.025
µ
S to 30 sec.
CCDLT1: 0.100mS
CCDLT2: 0.200mS
NOTE:
If you press the ENTER key, and the blinking data do not go to next, change
configuration setting Enter Data Next to YES.
Setting CR Values
To program the CR values for the selected channel, press the MODE,
^
, and ENTER
keys. The resistance values can be programmed in low voltage (CRL) or high voltage
(CRH) range. The current is always in high range. ALL resistance levels are
programmed in ohms. The slew rate is in A/
µ
S.
The following examples illustrate how to set CR values of Load module for model
MML-80-60-301.
1. Select
Range
Press MODE, and use the
^
and
v
key to select CRL, then press ENTER.
MODE SELECT
CRL
2. Set Resistor Level
There are 4000 discrete steps from 0 to full scale in each range. Set the main
resistor level1 (A) to 2 ohms by pressing 2, then ENTER. Set the level2 (B)
resistor level to 1 ohm by pressing 1, then ENTER.
CCL1: 2.000
Ω
CCL2:
1.000
Ω
3. Set Slew Rate
There are 250 discrete steps in each range. Set the rise and fall slew rates to 0.1
A/
µ
S by pressing ., 1, then ENTER for rise slew rate and ., 2, then ENTER for fall
slew rate.
CRL
:
0.10A/
µ
S
CRL
:
0.20A/
µ
S