Local Operation
4-7
Figure 4-3 CV Response Transfer Function (SLOW)
4.2.3 Setting the Program
The electronic Load is able to select customized basic tests, and link them into a
program test for automatic execution.
The PROG key is used to select a program, or recall a program for local control.
There are ten programs (1-10). Each program has ten sequences to map files from 1
to 100. Program 1 maps files from 1 through 10. Table 4-1 shows the relationship
between the program sequence and the corresponding file.
Program
1
Sequence
No. 1 2 3 4 5 6 7 8 9 10
Corresponding
File
No. 1 2 3 4 5 6 7 8 9 10
Program
2
Sequence
No. 1 2 3 4 5 6 7 8 9 10
Corresponding
File
No. 11 12 13 14 15 16 17 18 19 20
:
:
Program
10
Sequence
No. 1 2 3 4 5 6 7 8 9 10
Corresponding
File
No. 91 92 93 94 95 96 97 98 99 100
Table 4-1 Relationship Between Program Sequence and Corresponding File
When running a program, you must set its corresponding file parameters first. If one
program sequence is not enough for you to test the UUT, you can use the program
chain function to get more sequences.