Local Operation
4-11
of channel input reference level. The HIGH and LOW levels can be set by the value or
percentage selected in configuration
SPEC. ENTRY MODE.
The HIGH/LOW
percentage range is from 0 to 100%.
Press MODE, ENTER, SPEC to set the specifications of CC mode. Press 5, ENTER
to set CENTER level 5V.
VOLTAGE SPEC.
CENTER: 5.0000V
Press 5, ENTER to set HIGH level 5%.
VOLTAGE SPEC.
HIGH PCet: 5.0%
Press 5, ENTER to set LOW level 5%.
VOLTAGE SPEC.
LOW Pcet: 5.0%
The default setting of HIGH and LOW is 100%. The CENTER value is half of the
range. To select the specifications set by Value or Percentage, please refer to 4.1.6.
4.2.6 Setting the Configuration
The electronic Load provides useful features such as Von point, Current limit, and Sync
run. To use these powerful features, use the configuration setup procedures to set the
parameters relevant to your application needs. This procedure is only needed for
initial setup of a test operation. The configuration of each channel is stored
independently in the EEPROM of the Mainframe. To set configuration you must press
CONF.
Set the voltage range of CC mode
. There are two voltage ranges for CC mode.
High range is for high voltage and low range is for low voltage to get better voltage
resolution. The default setting of Vrange is HIGH.
CC Vrange Select
1:HIGH 2=LOW
Set Von point
. Von is the conduction voltage level when the electronic Load starts to
sink current and the UUT output reaches the Von voltage. The default setting of Von
voltage is 1V.
Von POINT
VOLTAGE: 3.50V