Sorensen MML Series Operation and Programming Manual
4-12
Set Von latch
. There are two operation modes for Von control. Von latch ON means
that Load will sink current continuously when Von voltage reaches. Von latch OFF
means that Load will stop sinking current when UUT voltage is under Von voltage.
The default setting of Von latch is OFF. Figures 4-4 and 4-5 show Von LATCH ON
and OFF current waveform separately.
Von LATCH
1:ON
2:OFF
CAUTION!
Do not set Von to 0V. This will cause an overshoot spike because the Load circuit
will be ON regardless of no UUT. If a UUT is applied, the overshoot may damage
the UUT regardless of the minimal current setting of the Load.
Figure 4-4 Von LATCH ON Current Waveform
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com