Sorensen MML Series Operation and Programming Manual
5-32
5.10.8 MODE Subsystem
:MODE
SP
?
<aard>
CCL
CCH
CCDL
CCDH
CRL
CRH
CV
MODE
Type: Channel-specific
Description:
This command sets operational modes of the electronic load.
Syntax:
MODE CCL
Sets CC mode of low range.
MODE CCH
Sets CC mode of high range.
MODE CCDL
Sets CC dynamic mode of low range.
MODE CCDH
Sets CC dynamic mode of high
range.
MODE CRL
Sets CR mode of low range.
MODE CRH
Sets CR mode of high range.
MODE CV
Sets CV mode.
Parameters:
CCL, CCH, CCDL, CCDH, CRL, CRH, CV
Example: MODE
CCL
Query Syntax:
MODE?
Returns the operational mode of the
electronic load.
Return Parameters: <aard>
Query Example:
MODE?
Return Example:
CCL
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com