Figure 5)
Mounting bracket installed on rear panel.
Figure 6)
Tilt mounting bracket up to insert or remove.
3. Remove the components from the packaging.
4. Place the lid on the canister and press it down firmly around entire perimeter
(see Figure 7).
Figure 7)
Canister lid assembled on the canister
Figure 8)
Overview of the canister lid components: P = Heater-cooler connection port,
V = vacuum port, T = tandem port (shown here with tandem port cap applied), S = pour
spout (shown here with pour spout cap applied), L = label for recording the installation
date. The hydrophobic filter is located inside the device below the vacuum port.
5. Apply the cap firmly over the matching pour spout (S) (see Figures 7 and 8).
6. Apply the tandem port cap tightly over the tandem port (T) (see Figures 7 and 8).
7. Place the suction canister with the mounted lid into the suction canister holder
(see Figures 9 and 10).
Figure 9)
Canister with lid inserted into the suction canister holder on side panel
Figure 10)
Canister with lid inserted into the suction canister holder on rear panel
Figure 11)
Overview of the
3T Aerosol Collection Set: 1 = suction canister, 2 = pre-
connected right angle connector, 3 = Heater-Cooler connection line, 4 = vacuum
source line
8. Attach the short 46 cm (1.5 ft) Heater-Cooler connection tube 3 (see Figure 11)
to the Heater-Cooler Overflow Outlet and then connect it to the port (P) of the
canister lid (see Figures 8, 9 and 10).
Caution: Do not connect the short 46 cm (1.5 ft) Heater-Cooler
connection tube to the center vacuum source port (V) (see Figure 8) as
this will potentially cause the hydrophobic filter to be exposed to water
and will limit the vacuum flow to the canister.
9. Attach the 366 cm (12 ft) 1/4” ID vacuum source line 4 with pre-connected right
angle connector 2 (see Figure 11) to vacuum port (V) (see Figures 8, 9 and 10),
making sure that the tubing is fully connected and no tubing is kinked.
10. If additional length is required to access the vacuum source, attach the 366 cm
(12 ft) 1/4” ID vacuum extension line with connector to the 366 cm (12 ft) ¼” ID
vacuum source line 4 (see Figure 11).
Note: If required, the vacuum extension line must be ordered
separately; REF 050900111.
11. Attach the other end of the 366 cm (12 ft) 1/4” ID vacuum source line 4 (see
Figure 11) to a regulated vacuum source (see Figure 12).
Caution: The vacuum source line must be connected to a vacuum
regulator to control and adjust flow to the 3T Aerosol Collection Set.
Do not exceed vacuum level of 640 mm Hg (25” Hg).