Service Manual: FUSION EQUINE DR® II |
51
3. Installing the FUSION EQUINE DR
®
II System
This chapter provides the information that you need to install the FUSION EQUINE DR
®
II system.
Tools Needed for Installation
The following basic tools are needed for installing the digital radiography system:
Digital volt-
ohm meter (20,000 Ω/V)
Basic hand tools including screwdrivers
Dosimeter
ESD wrist strap. Must be able to read uR per
exposure.
X-ray phantoms
Calibration filter supplied with panel or with the x-
ray system.
Charging the System Components
System components that use batteries are usually shipped with a minimal charge for
equipment and personnel safety reasons. Charge these components before system use.
Procedure
1.
Connect the AC-DC adaptor to the DC input on the tablet and to the power source. You
can operate the tablet while it charges. See
SOUND 14” TABLET controls, indicators,
on page 6 for more information.
2.
Charge the panel battery.
Charging the AX-B2735W panel battery
on page 51.
Charging the AX-B3543W detector batteries
on page 51.
Charging the AX-B2735W and AX-B3543W batteries
Note:
The use of accessories, batteries, battery chargers, or cables
other than those specified in this manual, with the exception of
those sold or provided by the manufacturer as replacement parts for
internal components, may result in increased emissions, decreased
immunity, or abnormal system operation. Use only the equipment and
accessories provided or specifically approved by the manufacturer.
About this task
Use the battery charger to charge the detector’s lithium-ion battery. The charger comes with a
power cable, which you connect to an AC power source. The charger has a rated input of 100
–
240 V AC (50/60 Hz, 0.7-0.37 A, 70-90 VA). The charger rated output is 12.33 V DC (1.2 A).
Summary of Contents for FUSION Equine DR II
Page 2: ... 2022 Sound company All rights reserved This page left blank intentionally ...
Page 10: ......
Page 35: ...Service Manual FUSION EQUINE DR II 17 1 System Overview ...
Page 38: ...20 Service Manual FUSION EQUINE DR II 1 System Overview Figure 10 Ready Indicator dimensions ...
Page 75: ...Service Manual FUSION EQUINE DR II 57 3 Installing the FUSION EQUINE DR II System ...