80 | Service Manual: FUSION EQUINE
DR®
4. Configuring the FUSION EQUINE DR
®
X-ray
System
Table 44: Advanced Configuration Options
Field
Details
High Watermark
Select the option that best fits the needs of the site. The options
are Never Auto Delete, 50, 75, and 90. Never Auto Delete is the
default value. Selecting any value other than Never Auto Delete
causes patient records to be deleted when the percent of used
hard disk space is greater than the value selected. The oldest
patient records are deleted first and records are deleted until the
percent of used disk space is less than the selected value.
Log off admin on
exit
If True is selected, and the Sound user is logged in to the
Window operating system, the Sound user is logged out
automatically when the software client is closed. True is the
default value. If False is selected, the Windows desktop is
displayed when the software client is closed.
Low Watermark
Select the option that best fits the needs of the site. The options
are 25, 50, and 75. 25 is the default value. Selecting any value
other than 25 causes patient records to be deleted when the
percent of used hard disk space is greater than the value
selected. The oldest patient records are deleted first and records
are deleted until the percent of used disk space is less than the
selected value.
Show Accession
Number
If set to True, the Accession Number field displays on Add
Patient, Edit Patient, and Add Study screens. If set to false, this
field does not appear on these screens.
Require Vet/Tech
Selection
Set this value to True to require the user to select a vet and/or
tech before closing the study. Options: True or False. Default:
False.
Require Image
Reject Reason
If set to True, a prompt will appear and you will have to select a
reason when rejecting an image. If set to false, this field does not
appear on these screens.
Gain Calibration
Interval
Specify the gain calibration interval from one of these options:
Off (calibration does not expire), Quarterly (calibration is valid for
three months), Semi-annual (calibration is valid for six months),
or Annual (calibration is valid for one year). The default value is
Off.
Acquisition
Session Interval
(hrs.)
Specify the acquisition session interval in hours. When a shot
is acquired into a study that already has a shot of the same
anatomy, the new shot is placed into a new series if the interval
between the first shot and the new shot exceeds the acquisition
session interval.
Summary of Contents for FUSION Equine DR II
Page 2: ... 2022 Sound company All rights reserved This page left blank intentionally ...
Page 10: ......
Page 35: ...Service Manual FUSION EQUINE DR II 17 1 System Overview ...
Page 38: ...20 Service Manual FUSION EQUINE DR II 1 System Overview Figure 10 Ready Indicator dimensions ...
Page 75: ...Service Manual FUSION EQUINE DR II 57 3 Installing the FUSION EQUINE DR II System ...
Page 77: ...Service Manual FUSION EQUINE DR II 59 3 Installing the FUSION EQUINE DR II System ...