74 | Service Manual: FUSION EQUINE
DR®
4. Configuring the FUSION EQUINE DR
®
X-ray
System
The digital radiography system is configured in the
Management
screen of the software application.
This chapter describes how to configure each part of the system.
Procedure
1.
Open the
Management
Displaying the Management screen
on page
64, for instructions.
2.
Configure Basic Options. See the topic,
on page 65, for
instructions.
3.
Configure Intermediate Options. See the topic,
Configuring Intermediate Options
on page
67, for instructions.
4.
Configure Advanced Options. See the topic,
on page 69,
for instructions.
5.
Configure the panel. See the topic,
on page 71, for instructions.
6.
Configure DICOM. See the topic,
on page 86, for information.
7.
Configure acquisition profiles. See the topic,
Configuring Acquisition Profiles
on page
97, for instructions.
Note:
For most sites, the default acquisition profiles are sufficient,
and no configuration is required.
8.
on page 106, for instructions.
9.
Configure logs. See the topic,
on page 144, for information about log file
options.
10.
Customize overlays. See the topic,
on page 116, for instructions.
11.
Select system backup options. See the topic,
Backing Up FUSION EQUINE DR Data
on page 122, for instructions.
Configuring the X-ray Generator
The x-ray generator is configured at the generator console.
Procedure
Review the documentation that accompanies the x-ray generator for instructions on
configuring the generator for use with the digital radiography system.
Important:
The x-ray generator exposure window must always be set to less than the
panel integration window (1 second).
Displaying the Management screen
The digital radiography system application is configured in the
Management
screen. Vet
Techs and Vets have some access to the
Management
screen, but Sound users have full
access.
Summary of Contents for FUSION Equine DR II
Page 2: ... 2022 Sound company All rights reserved This page left blank intentionally ...
Page 10: ......
Page 35: ...Service Manual FUSION EQUINE DR II 17 1 System Overview ...
Page 38: ...20 Service Manual FUSION EQUINE DR II 1 System Overview Figure 10 Ready Indicator dimensions ...
Page 75: ...Service Manual FUSION EQUINE DR II 57 3 Installing the FUSION EQUINE DR II System ...
Page 77: ...Service Manual FUSION EQUINE DR II 59 3 Installing the FUSION EQUINE DR II System ...