ST1952 Operation Manual
52
:VOLTage:AC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n
〉
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure AC voltage
Set the test speed and display digits
Query line cycle integration rate
Path to set measurement range
Select range (0 to 757.5)
Query range
Enable or disable auto range
Query auto range
specify reference (-757.5 to 757.5)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
SLOW
757.5
ON
0
OFF
Speed Commands
: NPLCycles <n>
Command Systax:
:VOLTage:AC:NPLCycles <n>
Set NPLC for ACV
:VOLTage:DC:NPLCycles <n>
Set NPLC for DCV
Command Parameter:
<n> = SLOW,FAST,PLAC4,PLAC5
Set the test speed and digit display
DEFault SLOW
SLOW Set the test speed as SLOW
FAST Set the test speed as FAST
PLAC4 Set the display digits as 4½
PLAC5 Set the display digits as 5½2
Query:
:NPLCycles?
Query the current test speed
Description:
The command is used for the basic measurement functions (except Frequency and Period) and test
speed (display digits).
:RANGe Commands
:[UPPer] <n>
Command syntax:
:VOLTage:AC:RANGe[:UPPer] <n>
Set measurement range for ACV
:VOLTage:DC:RANGe[:UPPer] <n>
Set measurement range for DCV