ST1952 Operation Manual
65
:ACQuire
Command syntax:
:FREQuency:REFerence:ACQuire
Acquire reference for FREQ
Description:
When one of these commands is sent, the measurement input signal is acquired and established as
the reference value. This command is typically used to zero the display. For example, if the
instrument is displaying a 10Hz offset, sending this command and enabling Reference zeroes the
display.
This command is functional only if the instrument is on the specified measurement function.
Sending this command while in any other function causes an error. Also, if the latest reading is
overflowed or a reading has not been triggered, an error occurs when this command is sent.
6.3.10 CAPacitance subsystem
Command
Function Description
Default
:CAPacitance
:RANGe
[:UPPer] <n
〉
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure capacitance
Path to configure measurement range
Range (1n,10n,100n,1u,10u,100u,1000u,10000u)
Query range
Enable or disable auto range
Query auto range
Specify reference (0~10000uF)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
1n
ON
0
OFF