ST1952 Operation Manual
45
6.3.3 FUNCtion2 subsystem
The commands in this subsystem are used to configure the second measurement function subsystems
and are summarized in Table 6-3.
Table 6-3
FUNCtion 2 Subsystem Commands Summary
Command
Command Description
:
FUNCtion 2
<
name
>
:
FUNCtion 2?
Select measurement function: ‘VOLTage:AC’, ‘VOLTage:DC’,
‘CURRent:AC’, ‘CURRent:DC’, ‘FREQuency’
Query the second function.
:FUNCtion2 Command
:FUNCtion2 <name>
Command syntax:
:FUNCtion2 <name>
Command Parameter:
<name> = ‘VOLTage:AC’
Select AC Voltage
‘VOLTage:DC’ Select DC Voltage
‘CURRent:AC’ Select AC Current
‘CURRent:DC’
Select DC Current
‘FREQuency’ Select Frequency
Query:
:FUNCtion2?
Query currently second programmed function
Description:
This command is used to select the second measurement function of the instrument.
6.3.4 UNIT subsystem
The UNIT subsystem is used to configure and control the measurement units for ACV, and DCV, and is
summarized in Table 6-4.
Table 6-4
UNIT Subsystem Commands Summary
Command
Command Description
Default
parameters
: UNIT
〈
name>
: UNIT?
Select measurement units: ‘V’, ’dB’, ‘dBm’
Query the measurement units
‘V’
UNIT command
:
UNIT
<name>
command syntax:
: UNIT <name>