ST1952 Operation Manual
68
the reference value. This command is typically used to zero the display. For example, if the
instrument is displaying a 0.1Ω offset, sending this command and enabling Reference zeroes the
display.
This command is functional only if the instrument is on the specified measurement function.
Sending this command while in any other function causes an error. Also, if the latest reading is
overflowed or a reading has not been triggered, an error occurs when this command is sent
6.3.11 HOLD subsystem
The commands in this subsystem are used to configure and control hold measurement function and are
summarized in Table 6-11.
Table 6-11
HOLD Subsystem Commands Summary
Command
Function Description
Default
:HOLD
:WINDow <NRf>
:WINDow?
:COUNt <NRf>
:COUNt?
:STATe <NRf>
:STATe?
Path to control Hold feature:
Set Hold window(%); 0.01 to 10
Query Hold window
Set Hold count; 2 to 100
Query HOLD count
Enable or disable HOLD
Query state of HOLD
1
10
OFF
:HOLD Command
The following commands are used to configure and control the HOLD feature.
:WINDow <NRf>
Command syntax:
:HOLD:WINDow <NRf>
Command Parameter:
<NRf> =
0.01 to 10
Set window (percent)
Query:
:WINDow?
Query Hold window
Description:
This command is used to set the window for HOLD. This window is expressed as a percent of the
“seed” reading for the Hold process.
:COUNt <NRf>
Command syntax:
:HOLD:COUNt <NRf>
Command Parameter:
<NRf> =
2 to 100
Specify HOLD count