43
Default measurement configuration for the CONFigure command
Measurement
parameters
Default Setting
AC
input
filter
(bandwidth)
20 Hz (medium speed filter)
Auto Zero
If the resolution setting causes NPLC to be less than 1, it will be turn OFF; if the
resolution setting causes NPLC to be greater than or equal to 1, it will be turn ON.
Range
AUTO (including voltage range for frequency and period measurement)
Number of samples
per trigger
1 sample
Trigger number
1 trigger
Trigger delay
Auto delay
Trigger source
Immediately
Trigger slope
NEGative
Mathematical function Forbidden. Other parameters have not changed.
Empty state of each
function
Forbidden
7.2.1.1 CONFigure?
Description:
Returns a quoted string indicating the current function, range, and resolution.
Syntax:
CONFigure?
Parameters:
(none)
Example:
conf?
Typical return:
DCI, 1.00000000E-04, 1.00000000E-09
7.2.1.2 CONFigure:VLOTage:DC
Description:
Configure the instrument for DC voltage measurement, setting all measurement parameters and
trigger parameters to their default values.
Syntax:
CONFigure[:VOLTage]:DC [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
<range>:{100m|1|10|100|1000}. Default: AUTO
Example:
CONF:VOLT:DC 100
7.2.1.3 CONFigure:VLOTage:AC
Description:
Configure the instrument for AC voltage measurement, setting all measurement parameters and
trigger parameters to default values
Syntax:
CONFigure[:VOLTage]:AC [{<range>|AUTO|MIN|MAX|DEF}
Parameters:
<range>:{100m|1|10|100|750}. Default: AUTO