44
Example:
CONF:VOLT:AC 100
7.2.1.4 CONFigure:CURRent:DC
Description:
Configure the instrument for DC current measurement, setting all measurement parameters and
trigger parameters to default values
Syntax:
CONFigure:CURRent:DC [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
<range>: {100 μA|1 mA|10 mA|100 mA|1 A|3 A|10 A}. Default: AUTO
Example:
CONF:CURR:DC 100m
7.2.1.5 CONFigure:CURRent:AC
Description:
Configure the instrument for AC current measurement, setting all measurement parameters and
trigger parameters to default values
Syntax:
CONFigure:CURRent:AC [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
<range>: {100 μA|1 mA|10 mA|100 mA|1 A|3 A|10 A}. Default: AUTO
Example:
CONF:CURR:AC 100m
7.2.1.6 CONFigure:CONTinuity
Description:
Configure the instrument for continuity measurement, setting all measurement parameters and
trigger parameters to default values
Syntax:
CONFigure:CONTiniuty
Parameters:
(none)
Example:
CONF:CONT
7.2.1.7 CONFigure:DIODe
Description:
Configure the instrument for diode measurement, setting all measurement parameters and trigger
parameters to default values
Syntax:
CONFigure:DIODe
Parameters:
(none)
Example:
CONF:DIOD
7.2.1.8 CONFigure:RESistance
Description:
Configure the instrument for two-wire resistance measurement, setting all measurement parameters
and trigger parameters to default values
Syntax:
CONFigure:RESistance [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
<range>: {10|100|1k|10k|100k|1MA|10MA|100MA} Default: AUTO
Example:
CONF:RES 10MA