8
DC resistance specification
[7] [8]
Range
Test
current
Resolution 24 hours
T
CAL
± 1°C
90 days
T
CAL
± 5°C
1 year
T
CAL
± 5°C
Temperature
coefficient /°C
10.00000Ω
10mA
10uΩ
0.0050+0.0040 0.0080+0.0060
0.0100+0.0080
0.0006+0.0008
100.0000 Ω
10mA
0.1mΩ
0.0030+0.0020 0.0080+0.0030
0.0100+ 0.0040
0.0006+0.0005
1.000000kΩ
1mA
1mΩ
0.0020+0.0005 0.0080+0.0010
0.0100+ 0.0010
0.0006+0.0001
10.00000kΩ
100uA
10mΩ
0.0020+0.0005 0.0080+0.0010
0.0100+ 0.0010
0.0006+0.0001
100.0000kΩ
10uA
100mΩ
0.0020+0.0005 0.0080+0.0010
0.0100+ 0.0010
0.0006+0.0001
1.000000MΩ 5uA
1Ω
0.0020+0.0010 0.0080+0.0010
0.0100+ 0.0010
0.0010+0.0002
10.00000MΩ 500nA
10Ω
0.0150+0.0010 0.0200+0.0010
0.0400+ 0.0010
0.0030+0.0004
100.0000MΩ 500nA/10M 100Ω
0.3000+0.0100 0.8000+0.0100
0.8000+ 0.0100
0.1500+0.0002
Continuity (conduction) test specification
Function
Range
Test current
24 hours
T
CAL
± 1°C
90 days
T
CAL
± 5°C
1 year
T
CAL
± 5°C
Temperature
coefficient /°C
Continuity
1kΩ
1mA
0.002 + 0.030
0.008 + 0.030
0.010+ 0.030
0.0010+ 0.0020
Diode test specification
Function
Range
Test current
24 hours
T
CAL
± 1°C
90 days
T
CAL
± 5°C
1 year
T
CAL
± 5°C
Temperature
coefficient /°C
Diode
[12]
5V
1mA
0.002 + 0.030
0.008 + 0.030
0.010+ 0.030
0.0010+ 0.0020
Temperature test specification
Temperature
PT100 (DIN/ IEC 751)
Probe ac 0.05°C
5 kΩ thermistor
Probe ac 0.10°C
Capacitance test specification
Range
24 hours
T
CAL
± 1°C
90 days
T
CAL
± 5°C
1 year
T
CAL
± 5°C
Temperature
coefficient /°C
1.0000nF
0.5+0.10
0.5+0.40
1.0+0.50
0.02+0.001
10.000nF
0.2+0.05
0.5+0.10
1.0+0.10
0.02+0.001
100.00nF
0.2+0.05
0.5+0.10
1.0+0.10
0.02+0.001
1.0000uF
0.2+0.05
0.5+0.05
1.0+0.10
0.02+0.001
10.000uF
0.2+0.05
0.5+0.05
1.0+0.10
0.02+0.001
100.00uF
0.2+0.05
0.5+0.05
1.0+0.10
0.02+0.001
1.0000mF
0.2+0.05
0.5+0.05
1.0+0.20
0.02+0.001
10.000mF
0.5+0.20
0.5+0.30
1.0+0.50
0.02+0.001
AC voltage test specification
[6] [9] [10]
Frequency/range
24 hours
T
CAL
± 1°C
90 days
T
CAL
± 5°C
1 year
T
CAL
± 5°C
Temperature
coefficient /°C
3 Hz - 5 Hz
1.00+0.02
1.00+0.02
1.00+0.03
1.00+0.03