2.2 Supplementary specifications
Display readings and measuring speed
DC resistance, DC voltage, DC current
[1]
Integral time
Default
digits
[2]
Reading/second
[3]
System
speed
[4]
Extra error
100PLC/2s(1.67s)
0.5(0.6)
0.5(0.6)
0
10PLC/200ms(167ms)
5(6)
5(6)
0
1PLC/20ms(16.7ms)
45(55)
45(55)
0.001% range
0.2PLC/4ms
200
50
0.001% range
[5]
0.02PLC/400μs
1000
50
0.01% range
[5]
AC voltage, AC current
[2] [3]
Full scale display
Default display
Reading / second
Speed
1199999
119999
7sec./reading
[6]
Slow
1199999
119999
1
[6]
Medium
1199999
119999
1.6
[6]
Fast
1199999
119999
10
Fast
[7]
1199999
119999
50
Fast
[8]
[1]. Reading rate at 60-Hz and 50-Hz operating conditions, turning off auto zero, within a fixed range.
[2]. The number of display digits in the case of automatically displaying the number of digits. If a fixed number of
digits are selected, the number of displayed digits does not change due to the test speed.
[3]. The average speed at which the ADC continuously sampling to the data transmission and display.
[4]. Read data by instructions.
[5]. Add 20μV to DCV; increase resistance by 20 mΩ. When DCA increases by 0.2μA, 100uA, 10mA range, the
range error is increased by 10 times.
[6]. Read stable data. The actual ADC sampling speed is faster than the stable speed.
[7]. External trigger or remote operation, using the default stable delay (automatic delay).
[8]. Use the default maximum value, set the delay to expire.