45
7.2.1.9 CONFigure:FRESistance
Description:
Configure the instrument for four-wire resistance measurement, setting all measurement parameters
and trigger parameters to default values
Syntax:
CONFigure:FRESistance [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
<range>: {10|100|1k|10k|100k|1MA|10MA|100MA} Default: AUTO
Example:
CONF: FRES 10MA
7.2.1.10 CONFigure:FREQuency
Description:
Configure the instrument for frequency measurement, set all measurement parameters and trigger
parameters to default values
Syntax:
CONFigure:FREQuency [{<range>|MIN|MAX|DEF}]
Parameters:
<range>: {100m|1|10|100|750} Default: AUTO
Example:
CONF:FREQ 10
7.2.1.11 CONFigure:PERiod
Description:
Configure the instrument for periodic measurements, setting all measurement parameters and trigger
parameters to default values
Syntax:
CONFigure:PERiod [{<range>|MIN|MAX|DEF}]
Parameters:
<range>: {100m|1|10|100|750} Default: AUTO
Example:
CONF:PER 10MA
7.2.1.12 CONFigure:TEMPerature
Description:
Configure the instrument for temperature measurement and set all measurement parameters and
trigger parameters to default values
Syntax:
CONFigure:TEMPerature [{FRTD|RTD|FTHermistor|THERmistor}]
Parameters:
<probe_type>: FRTD|RTD|FTHermistor|THERmistor
Example:
CONF:TEMP RTD
7.2.1.13 CONFigure:CAPacitance
Description:
Configure the instrument for capacitance measurement, setting all measurement parameters and
trigger parameters to default values
Syntax:
CONFigure:CAPacitance [{<range>|AUTO|MIN|MAX|DEF}]
Parameters:
<range>:{1n|10n|100n|1u||10u|100u|1m|10m} Default: AUTO
Example:
CONF:CAP 10u
7.2.2 MEASure subsystem
With MEASure? queries, you can use a single command to select functions and ranges. All other parameters are