56
7.2.5.8 [SENSe:]CURRent:DC:SECondary
Description:
Select the auxiliary measurement function for the DC current measurement.
Please use DATA2? to read auxiliary measurement results.
Syntax:
[SENSe:]CURRent:DC:SECondary {OFF|BEForemath }
[SENSe:]CURRent:DC:SECondary?
Parameters:
{OFF|BEForemath}
Example:
CURR: DC: SEC BEF
7.2.5.9 [SENSe:]CURRent[:DC]:ZERO:AUTO
Description:
Enable or disable the auto zero mode for DC current.
ON: The DMM internally measures the offset after each measurement. This measurement is then subtracted from
the previous reading. This prevents the offset voltage on the DMM input circuit from affecting the measurement
accuracy.
OFF: The instrument uses the last measured zero return measurement and subtracts it from each measurement. It
reads a new zero reading whenever you change the function, range or integration time.
Syntax:
[SENSe:] CURRent [:DC]:ZERO:AUTO {ON|OFF}
[SENSe:] CURRent [:DC]:ZERO:AUTO?
Parameters:
{ON|1|OFF|0}
Example:
CURR: ZERO: AUTO ON
7.2.6 [SENSe:]{RESistance|FRESistance} subsystem
7.2.6.1 [SENSe:]{RESistance|FRESistance}:RANGe
Description:
Set a fixed range for 2-wire or 4-wire resistors
Syntax:
[SENSe:] {RESistance|FRESistance}:RANGe {<range>|MIN|MAX}
[SENSe:] {RESistance|FRESistance}: RANGe? [{MIN|MAX|DEF}]
Parameters:
<range>:{10|100|1k|10k|100k|1MA|10MA|100MA }
Example:
RES: RANG 10MA
7.2.6.2 [SENSe:]{RESistance|FRESistance}:RANGe:AUTO
Description:
Enable or disable the autoranging for 2-wire or 4-wire resistance measurements
The autoranging range can be adjusted down to less than 10% of the range and can be adjusted upwards to more
than 120% of the range.
Syntax:
[SENSe:] {RESistance|FRESistance}:RANGe:AUTO {ON}OFF}
[SENSe:] {RESistance|FRESistance}: RANGe: AUTO?
Parameters:
{ON|1|OFF|0}
Example:
RES: RANG: AUTO ON