ST2826 Operation manual
Ver1.0
◇
102
<high limit n>: NR1,NR2 or NR3 format
,
the high limit data in the line n.
e.g.
:
WrtCmd(―LIST:BAND1 A,10,20‖)
WrtCmd(―LIST:BAND3 OFF‖)
query syntax
:
LIST:BAND<n>?
return
:
<parameter>,<low limit n>,<high limit n>
8.1.11APERture subsystem command set
APERtures
subsystem command set is used to set the measurement speed, the average times.
Char ‗?‘ can be used to query the current speed, the average times.
command syntax
:
FAST
APERture
MEDium [,<value>]
SLOW
FAST
:
fast, 30times/s
MEDium
:
medium, 10 times/s
SLOW
:
slow, 2 times/s
<value>
1 to 128(NR1 format), the average times.
e.g.
:
WrtCmd(―APER MED,55‖);
query syntax
:
APERture?
return
:
FAST
MED ,<NR1><NL^END>
SLOW
8.1.12TRIGger subsystem command set
TRIGger
subsystem command set is used to set the trigger source, the delay time after the
trigger and trigger instrument test.
Command hierarchical chart:
TRIGger
[:IMMediate]
:SOURce INTernal
Summary of Contents for ST2826